Vue de la page 59
4-4 QX Hawk Industrial Imager User Manual
Video
Select Lens (Standard QX Hawk Only)
If you change the QX Hawk’s lens, you must configure the Lens Type parameter to match
the new lens type.
Click the Select Lens button on the Video tab.
To verify the lens type of your camera, power-off the QX Hawk (do not look into the targeting
laser) and look at the outer rim of the unit’s camera lens. You will see one of the following:
’ and ‘15’ = Lens Type 1
●●’ and ‘30’ = Lens Type 2
●●●’ and ‘45’ = Lens Type 3 (Default)
’ and ‘12’ = Lens Type 4
Select your lens type in the Select Reader Optics dialog and click OK.
You can calibrate the new lens by using the <@OPTIC> command.
Important: <@OPTIC> is not supported by the QX Hawk C-Mount Imager.
This parameter is “sticky”. Once saved with a <Z> or <Zc> command, it can only be
defaulted with <Zrdall> or <K525d>.
30 Degree Lens 45 Degree Lens
15 Degree Lens12 Degree Lens