Optoelectric Database Modes
8-6 MS-880 Industrial Long Range Scanner User’s Manual
8–Optoelectrics
Timer Mode
Calibration with Multilabel
With Multilabel enabled for more than one label, and either Timer or Automatic/
Static mode enabled, the scanner may not locate labels that are scanned at various
ranges. The reason for this is that once a label is located from the optical database
index, the scanner will continue to search for labels that meet that index of focus,
tracking and gain settings.
Automatic/Static Mode
Definition: When this mode is selected, the scanner cycles through database set-
tings at a time interval determined in Timer Value setting. (See “Timer
Value” on page 8-11.)
Important Note: The maximum time to move from the most inside
focus position to the most outside focus position is 400 ms. One focus
position change takes 10 ms.
Usage: Useful in applications in which there is enough time to cycle through the
database index.
Serial Cmd: <K226,mode,# active database settings,latch on trigger,timer value,
control trigger 1,control trigger 2,control trigger 3>
Definition: Selects the optimum setting from the Configuration Database.
When selected, the scanner cycles once through all the Configuration
Database settings from the closest to the furthermost focal position,
testing first for optimal signal, then for a successful decode.
Note: Since focus positions are tested in ascending order, the closer
the symbols are to the scanner, the quicker the best focus position will
be found.
Important Note: The maximum time to move from the most inside
focus position to the most outside focus position is 400 ms. One focus
position change takes 10 ms.
Usage: Useful in a static application in which the bar code symbol is in front of
the scanner when the read cycle begins.
ESP:
Serial Cmd: <K226,mode,# active database settings,latch on trigger,timer value,
control trigger 1,control trigger 2,control trigger 3>