
Laser Setup
9-8 MS-880 Industrial Long Range Scanner User’s Manual
9–Raster/Laser Con-9–Raster/Laser Con-
Laser Off Time
When setting up Laser Framing in ESP-MP, use the Off Time and On Time arrows to
constrain or expand the laser framing area.
Laser On Time
Laser Power
Definition: Duration of laser off time, in increments of 1/255th of the total scan
width.
Important: The combined values of Laser Off Time and Laser On
Time cannot exceed 255 which is the duration of one scan.
Serial Cmd: <K700,laser on/off status,laser framing status,laser off time,laser on
time,laser power>
Default: 0
Options: 0 to 255
Definition: Duration of laser on time, in increments of 1/255th of the total scan
width.
Important: The combined values of Laser Off Time and Laser On
Time cannot exceed 255 which is the duration of one scan.
Serial Cmd: <K700,laser on/off status,laser framing status,laser off time,laser on
time,laser power>
Default: 255
Options: 0 to 255
Definition: The user can select the Low or High laser power.
Usage: Applications that read past 45 inches typically use High.
Serial Cmd: <K700,laser on/off status,laser framing status,laser off time,laser on
time,laser power>
Default: Low
Options: 0 = Low (laser power = 2 mW)
1 = High (laser power = 3.25 mW)
Commentaires sur ces manuels