Autocalibration by ESP
6-4 MS-880 Industrial Long Range Scanner User’s Manual
6–Autocalibration
Autocalibration by ESP
1. Click on the Autocalibration button on the overhead toolbar.
2. Click on the Auto Cal button to start autocalibration.
An Autocalibration in
Progress message will be
displayed on your monitor
along with the pass number
and information on the focus,
gain, and tracking.
Allow some time for the scan-
ner to cycle through the
focus, gain, and tracking set-
tings. Do not move or disturb
scanner or label while Auto-
calibration is in progress. (It
might take 2 to 3 minutes to
complete.)
3. When completed, the settings
that provided the highest
read rate will be selected. You
will see one of the following
messages: Autocalibration
Passed, Autocalibration
Failed, or Original Set-
tings Were Optimal.
If it passes, you will con-
tinue to see the read rate
percentage and label data in
the window below the scan-
ner image.
Note that in the figure 6-4
the focus position remains
the same as it was in figure
6-3, but the Gain and
Tracking have been fine-
tuned during the second
pass.
The graph in figure 6-4 shows
the results of this procedure.
Figure 6-3 Autocalibration In Process
Figure 6-4 Autocalibration Passed